Calibrate and deskew an acquisition, optionally deconvolving it. Saves uint16 output beside the acquisition by default.
With --deconvolve, 3D scan refinement is inferred from recorded spacing toward
0.4 micrometers. For example, a 1.2 micrometer acquisition uses factor 3. Already
finer acquisitions retain their spacing; coarser steps must be integer multiples
of 0.4 micrometers.
Measured PSFs must be sampled at the inferred reconstruction spacing.
| Option | Description |
|---|---|
--deconvolve |
Apply RLGC deconvolution before deskewing. |
--flatfield-correction |
Include illumination in the deconvolution measurement model, or divide calibrated data by it when deconvolution is disabled. |
--save-float32 |
Preserve fractional intensities in float32 output. |
--skip-empty-below VALUE |
Zero channel tiles with insufficient signal above this intensity. |
--skip-empty-min-signal-fraction VALUE |
Required fraction above the cutoff; default 0.01. |
--resume |
Continue compatible completed tiles. If omitted, partial outputs offer Resume / Overwrite / Cancel (default: Cancel); completed outputs are overwritten. |
--output PATH |
Set the output directory. |
--live PATH |
Process during acquisition using a supplied CYX illumination TIFF; pass the acquisition directory. |
--live-wait |
Wait for successful live completion and all raw tiles, then process offline. Combine with --flatfield-correction to estimate missing illumination. Mutually exclusive with --live; pass the acquisition directory. |
--time-range START STOP |
Select timepoints. |
--pos-range START STOP |
Select positions. |
--z-downsample-level VALUE |
Deskewed Z reduction factor; default 2. |
--crop-after-deskew |
Crop Y to the region supported across all Z planes. |
--no-max-projection |
Disable per-tile and stage-placed maximum-Z images. |
--no-create-fused-max-projection |
Disable the stage-placed maximum-Z mosaic. |
--write-fused-max-projection-tiff |
Also save that mosaic as OME-TIFF. |
--decon-crop-scan VALUE |
Set retained reconstruction scan planes per deconvolution chunk; default automatic. |
--decon-gpu-id VALUE |
Select the CUDA device; default 0. |
--decon-verbose VALUE |
Opt-in solver diagnostics: default 0 (off); 1 reports chunk geometry, iterations and termination reason. |
--decon-fallback-step-scan VALUE |
Set chunk reduction after GPU allocation failure. |
--decon-psf-paths PATH |
Supply channel-ordered PSFs; repeat for each channel. |
--eager-mode |
Retained for compatibility; has no effect on current KL stopping. |
uv run process "/path/to/acquisition" --deconvolve --flatfield-correctionStart before acquisition finishes, without a precomputed illumination file:
uv run process "/path/to/acquisition-directory" --live-wait --flatfield-correction --deconvolveSee the live acquisition guide for required controller sidecars and completion behavior.
Append --help to list all options and defaults.
Nonempty deconvolution chunks terminate through KL divergence by default: both prediction-to-half divergences must increase, and the preceding reconstruction is restored. Fraction/change thresholds do not preempt KL. All-zero calibrated chunks return zero without iterations. Each overlapping scan chunk reaches its first local KL stop and retains its iteration state in system memory. Chunks then continue to the latest first local KL stop across the volume, without repeating completed iterations.
Sparse reconstruction uses zero boundaries with detector Y/X padding to prevent
edge wraparound. This padding remains enabled even if the array solver's native
pad_yx compatibility setting is disabled.
The photon-GC solver revision is recorded with the processing configuration. Resume rejects incompatible solver configurations; start a fresh output or intentionally overwrite instead. Processing without deconvolution retains its existing calibration and flatfield behavior.